Toolyard

14,000 × g to RPM

14,000 × g at common rotor radii

Rotor radiusSpeed to set
5 cm15,826 RPM
6 cm14,447 RPM
7 cm13,375 RPM
8 cm12,511 RPM
8.5 cm12,138 RPM
9 cm11,796 RPM
10 cm11,190 RPM
12 cm10,215 RPM
15 cm9,137 RPM
20 cm7,913 RPM

How to set a centrifuge to 14,000 × g

  1. Find the radius of your rotor in cm, from the centre of the rotor to the bottom of the tube. It is printed on the rotor or in its manual. Enter it above; 14,000 × g is already filled in.
  2. The RPM to set is shown as you type. The table below gives it for common radii.
  3. Set the centrifuge to that RPM, or enter the × g directly if it has an RCF mode.

Frequently asked questions

How many RPM is 14,000 × g?

It depends on the rotor. At an 8.5 cm radius, typical for a microcentrifuge, 14,000 × g is about 12,138 RPM; at 10 cm it is about 11,190 RPM. The formula is RPM = √(RCF ÷ (1.118 × 10⁻⁵ × r)), with r in cm.

Is 14,000 g the same as 14,000 RCF?

Yes. RCF, relative centrifugal force, is written as a multiple of g, so 14,000 × g, 14,000 RCF and 14,000 g all mean the same force.

Why does a bigger rotor need fewer RPM?

Force grows in proportion to the radius, so a rotor twice the size reaches 14,000 × g at 8,583 RPM of the speed (1 ÷ √2, about 71%).

Common centrifuge speeds

More biology lab calculators